AJSM Click here for details!
HOME HELP CONTACT US SUBSCRIPTIONS ARCHIVE SEARCH
 QUICK SEARCH:   [advanced]


     

Sign In to gain access to subscriptions and/or personal tools.
First published on July 1, 2008, doi:10.1177/0363546508319314

(American Journal of Sports Medicine 2008;36:2179.)

A more recent version of this article appeared on November 1, 2008
This Article
Right arrow Full Text (AJSM PreView[PDF])
Right arrow All Versions of this Article:
36/11/2179    most recent
0363546508319314v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowRequest Permissions
Right arrow Request Reprints
Google Scholar
Right arrow Articles by Goleski, P.
Right arrow Articles by Pearle, A. D.
PubMed
Right arrow PubMed Citation
Right arrow Articles by Goleski, P.
Right arrow Articles by Pearle, A. D.

Article

Reliability of Navigated Lower Limb Alignment in High Tibial Osteotomies

Patrick Goleski1, Blaine Warkentine, MD2, Darrick Lo3, Cassie Gyuricza, MD3, Daniel Kendoff, MD4, Andrew D. Pearle, MD3*

1 Hospital for Special Surgery and University of Michigan Medical School
2 Hospital for Special Surgery and BrainLAB USA
3 Hospital for Special Surgery
4 Hospital for Special Surgery and Hannover Medical School

* To whom correspondence should be addressed. E-mail: PearleA{at}hss.edu.


   Abstract

Background: Navigation allows for determination of the mechanical axis of the lower extremity during high tibial osteotomy (HTO) procedures. The objectives of this study were to (1) evaluate the reliability of noninvasive registration with an image-free navigation system for HTO and (2) determine the accuracy of the navigation system to monitor changes in lower limb alignment as compared with alignment measured with a novel 3-dimensional computed tomography method.

Hypothesis: Navigated limb alignment demonstrates good reliability and accuracy in all 3 planes.

Study Design: Descriptive laboratory study.

Methods: Thirteen cadaver legs were used to examine the intra- and interobserver registration reliability of 3 observers. Initial coronal, sagittal, and axial alignment was measured on 6 legs, 3 times each, at intervals >36 hours. Navigated HTOs were then performed on all 13 legs, pre- and postoperative alignment was recorded, and data were compared with equivalent measures obtained by 3-dimensional computed tomography. Reliability and accuracy data were both analyzed using intraclass correlation coefficients with the following established thresholds: good, >0.75; fair, 0.4 to 0.75; and poor, <0.4.

Results: Intraclass correlation coefficients for intraobserver reliability were categorized as follows: varus-valgus, good; flexion-extension, fair; and femoral-tibial rotation, poor. For interobserver reliability, results were varus-valgus, fair; flexion-extension, fair; and femoral-tibial rotation, poor. Intraclass correlation coefficients for navigation accuracy were varus-valgus, good; tibial slope, good; and tibial torsion, poor. Maximum differences in navigation–computed tomography measurements were varus-valgus angle, 4.5°; tibial slope, 8.8°; and tibial torsion, 16.5°.

Conclusion and Clinical Relevance: Navigation may be reliable and clinically useful for dynamic monitoring of coronal leg alignment but has limits in determination of sagittal and axial plane alignment.